Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version Earlier than 03.00)
TMS60 Operator’s Manual
12 - 11
3. Select
Exit
to save the settings and close the
Waveform Setup
menu.
■
When
Analysis Lead Setup
is set to
Off
, the displayed leads and analysis leads
should be configured respectively.
Follow the steps below to configure the ECG waveform setting:
1. In the
HR
section of the
ECG
tab, select
Waveform Setup
.
The
Waveform Setup
menu displays.
2. Select the options described in the following table to configure the ECG wave
forms.
Options
Description
Settings*
Lead
Waveform size
ECG 1
Select the desired ECG lead and set the
corresponding gain for ECG 1.
I,
II
, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, × 4
ECG 2
Select the desired ECG lead and set the
corresponding gain for ECG 2.
I
, II, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, × 4
ECG 3
Select the desired ECG lead and set the
corresponding gain for ECG 3.
I, II, III, aVR, aVL,
aVF,
V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, and × 4
*
The factory default settings are in bold.
Options
Description
Settings*
Lead
Waveform size
ECG 1
Select the desired ECG lead and set the
corresponding gain for ECG 1 to display
in the waveform area.
I,
II
, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, × 4
ECG 2
Select the desired ECG lead and set the
corresponding gain for ECG 2 to display
in the waveform area.
I
, II, III, aVR, aVL,
aVF, V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, × 4
ECG 3
Select the desired ECG lead and set the
corresponding gain for ECG 3 to display
in the waveform area.
I, II, III, aVR, aVL,
aVF,
V
× 0.125, × 0.25, × 0.5,
× 1
, × 2, × 4
Primary lead
Configures the primary analysis lead.
I,
II
, III, aVR, aVL,
aVF, V
None
Secondary lead
Configures the secondary analysis lead.
I
, II, III, aVR, aVL,
aVF, V
None
*
The factory default settings are in bold.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...