General Problems
Troubleshooting
14 - 2
TMS60 Operator’s Manual
14.1
General Problems
The following table lists the problems that are likely to occur. If the problem persists after
corrective actions have been taken, contact your service personnel.
Symptom
Possible cause
Solution
No RF Signal (The RC60
does not receive valid
data for consecutive 5
seconds)
1.
The TD60’s battery charge is to
be depleted.
2.
The patient is out of the antenna
array coverage area.
3.
The TD60 is not connected with
the ECG leadwire.
1.
Check if the TD60 battery charge
is depleted.
2.
Check if the patient is out of the
coverage area.
3.
Check if the TD60 is connected
with the ECG leadwire properly.
RF Interference (The
TD60 receives three
consecutive wrong
frames.)
The TD60 signal is weak.
Check if the Patient is at the edge of the
coverage area or in an elevator. Remove
behind a reinforced concrete wall.
SpO
2
No Pulse
The SpO
2
sensor failed to obtain pulse
signal.
Check the patient’s condition and
change the sensor application site. If
the error persists, replace the sensor.
The SpO2 data is not
displayed on the
CentralStation.
1.
The SpO
2
module is not con-
nected to the TD60.
2.
There may be error in the SpO
2
module.
1.
Connect the SpO
2
module to the
TD60.
2.
If there is an error in the SpO
2
module, replace the SpO
2
mod-
ule with a new one.
ECG noise
ECG waveforms are overlapped with the
noise interference.
Check if the ECG leadwire is intertwined
with cables of other devices.
ECG Signal Saturated
The TD60 detects that ECG signal
saturation or overload.
1.
Check if patient has excessive
movement.
2.
Check if the electrodes are in
good contact with the skin.
3.
Check if the electrodes operat-
ing time is over the electrode
service life.
Wrong Channel
The configured frequency of the
transmitter does not match the
channel.
Reconfigure the frequency of the
transmitter to match the channel.
The AC power
indicator on the RC60
is off.
1.
The RC60 is not connected to the
power.
2.
The RC60 is power off.
1.
Check if the RC60 is connected
to the power.
2.
Check if the RC60 sounds a beep
when turn the receiver on.
The TD60 or SpO
2
module restarts
repeatedly.
The TD60’s battery charge is to be
depleted.
Replace with new batteries.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...