Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 18
TMS60 Operator’s Manual
12.2.4
ST Monitoring
A normal ECG waveform (for ST monitoring)
The ST segment of an ECG waveform (as shown in the above figure) represents the
period from the end of ventricular de-polarization, to the beginning of ventricular re-
polarization, or the end of the QRS complex (the J point) and the beginning of the T-
wave. ST Segment analysis is used to monitor the oxygen supply and the viability of the
heart muscle.
ST deviation is the vertical distance between the isoelectric (ISO) point level and signal
level at ST point.
The ISO point is located between the end of the P-wave and the onset of the QRS com-
plex. The ISO point provides the baseline for this measurement.
The ST point is a fixed distance from the J point at the end of the QRS complex. The ST
point can be configured to 40, 60, or 80 milliseconds past the J-point, independent of
the heart rate. By default, the ST point is positioned as follows:
■
At 80 milliseconds for heart rates less than or equal to 120 beats per minute
■
At 60 milliseconds for heart rates greater than 120 beats per minute
All available ECG leads are analyzed to measure deviations in the ST segment.
Selecting leads that contain the least amount of baseline flutter will improve measure-
ment accuracy, but accurate ST deviation measurement is dependent on the correct
location of the ISO and ST points.
■
ST segment analysis calculates ST segment elevations and depressions for
individual leads and then displays them as numerics in the ST digital area.
■
A positive value indicates ST segment elevation; a negative value indicates ST
segment depression.
R
P
T
S
ST P
o
in
t
J P
o
in
t
IS
O
P
o
int
40 to 80 msec
ST deviation
Depression or Elevation
ST Segment
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...