12.2.5.3
Arrhythmia Settings
The following table lists the arrhythmia settings in the
Arrhythmia Analysis
section of
the
ECG
tab.
12.2.5.4
Relearning
A relearn can be done for arrhythmia, ST analysis, or for both simultaneously.
The TD60 initiates the learning process for ST or Arrhythmia analysis after any of the fol-
lowing:
■
TD60 power-up
■
Return to normal monitoring from the Standby mode
■
Enabling ST or Arrhythmia analysis
■
The lead has been changed in ECG 1 waveform (3 lead only)
■
Patient size changes
■
Selecting the
Relearn
button in the
Arrhythmia Analysis
section
Selecting the the
Relearn
button in the
Arrhythmia Analysis
section is recommended
after one or more of the following:
■
ECG electrodes have been repositioned
■
Eight hours have passed since the last relearn
■
Significant changes occurred to the patient QRS complex
■
Significant changes occurred to the patient ECG rhythm
Options or Buttons
Description
Settings*
Arrhythmia Alarms
Configures the arrhythmia alarm settings.
Alarm Settings" on page 12 - 6
for
details.
Relearn
Enables an
arrhythmia relearning.
for details.
QRS Threshold Settings
Configures the QRS threshold.
for details.
NOTE
•
ST Analysis must be turned on before it will relearn.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...