EMC
TMS60 Operator’s Manual
B - 3
If the system is operated within the electromagnetic environment listed in Table Guid-
ance and declaration - electromagnetic immunity, the system will remain safe and pro-
vide the following essential performance:
■
Operating mode
■
Accuracy
■
Function
■
Accessories identification
■
Data stored
RF emissions
CISPR 11
Class A
The system is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies build-
ings used for domestic purposes.
Harmonic
Emissions IEC61000-3-2
Class A
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Complies
Note: The TD60 Transmitter and BP10 (subpart) are classified by is their using environment or intended func-
tion.
NOTE
•
The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information pro-
vided below.
•
Other devices may affect this device even though they meet the require-
ments of CISPR.
•
When the inputted signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
•
The EMISSIONS characteristics of this device make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residen-
tial environment (for which CISPR 11 class B is normally required) this
device might not offer adequate protection to radio-frequency commu-
nication services. The user might need to take mitigation measures,
such as relocating or re-orienting the device.
•
If the essential performance is lost or degraded, it may be necessary to
take mitigation measures, such as re-orienting or relocating the ME
EQUIPMENT or ME SYSTEM or shielding the location or stopping using
the monitor and contact the service personnel.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...