Safety
Safety Information
TMS60 Operator’s Manual
1 - 5
1.1.2
Cautions
CAUTION
•
Do not let the display of the TD60 directly touch the patient when the
display is on.
•
When the Central Monitoring System presents the alarm “RF No Signal”,
the settings being performed on the TD60 may not be transferred to the
central monitoring system. Check the patient condition and the settings
on the Central Monitoring System.
•
When disposing of the packaging material, be sure to observe the appli-
cable waste control regulations and keep it out of children’s reach.
•
After the configurations, such as the patient category, paced status, are
changed at the TD60, the medical personnel should check those config-
urations at the Central Monitoring System to make sure both of the con-
figurations are consistent.
•
Mindray takes no responsibility for controlling the radio frequency envi-
ronment in a hospital. If interference for the operating frequency of
telemetry equipment exists, the telemetry equipment performance will
be affected. Exercise caution when selecting the operating frequency of
all the wireless equipment in a hospital as this is very important to avoid
mutual interference among them.
•
Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason make sure that all exter-
nal equipment operated in the vicinity of the equipment comply with
the relevant EMC requirements. Mobile phone, X-ray equipment or MRI
equipment are a possible source of interference as they may emit higher
levels of electromagnetic radiation.
•
Always install or carry the equipment properly to avoid damage caused
by drop, impact, strong vibration or other mechanical force.
•
Dry the equipment immediately in case of rain or water spray.
•
The system generates and uses the Radio Frequency (RF) energy. If it is
not installed correctly or not used as per the manual, RF interference to
other equipment could result.
•
When the patient carrying a TD60 moves in a hospital, signal quality can
be impacted on an ambulatory patient by the construction materials
used within the hospital..
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...