Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 10
TMS60 Operator’s Manual
12.2.2.2
Waveform Setup
The ECG leads waveforms displayed in the waveform area are defined as the displayed
lead.
The CMS uses information from two leads to detect beats and to compute HR. These two
leads are referred to as the primary and secondary leads. The user can select any of the
available leads (depending on whether a 3 or 5 lead cable is used) as primary, secondary
or other analysis leads.
For best results, the following guidelines should be used:
■
The QRS complex should be tall, narrow and preferably either completely above or
below the baseline. If at all possible, avoid selecting a lead where the QRS complex
is biphasic.
■
The P-waves and T-waves should be small compared to the QRS. They should be
less than about 0.2 mV.
Changing the Setting for Analysis Leads and Displayed Leads
The default system setting for the analysis leads and displayed leads is
On
, which means
that the analysis leads are consistent with the displayed leads. You can change the
default setting as following steps, if necessary.
1.
At the CMS, select
Admin Setup
→
input the password
→
select
OK
.
2.
In the
Admin Setup
menu, select the
Telemetry
tab.
3.
On the left side of the
Telemetry
tab, select
Analysis Lead Setup
.
The corresponding setting displays to the right of
Telemetry
.
4.
Set
Consistent with Displayed Lead
to
On
or
Off
.
◆
On
: the analysis lead is the same as the displayed lead.
◆
Off
: the analysis lead is different from the displayed lead.
Changing ECG Wave Settings
In the
Waveform Setup
menu, you can configure the displayed leads and analysis leads
as desired. The lead settings are dependent on the setting of
Analysis Lead Setup
.
■
When
Analysis Lead Setup
is set to
On
, the ECG waveform configurations are
applied for the displayed leads and analysis leads.
Follow the steps below to configure the ECG waveform setting:
1.
In the
HR
section of the
ECG
tab, select
Waveform Setup
.
The
Waveform Setup
menu displays.
2.
Select the options described in the following table to configure the ECG
waveforms.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...