Monitoring with the TMS60 at the CMS
Monitoring at the CMS (Version 03.00 or Above)
TMS60 Operator’s Manual
12 - 47
PVCs/min
PVCs/min exceeds high limit.
Nonlethal
arrhythmia
R on T
R on T detected in normal heartbeats.
Run PVCs
More than 2 consecutive PVCs.
Couplet
Paired PVCs detected in normal heartbeats.
PVC
One PVC detected in normal heartbeats.
Vent Rhythm
The consecutive PVCs ≥ the V-Brady PVCs limit, and the HR ≥ V-
Brady Rate limit but < the V-Tach Rate limit.
Bigeminy
A dominant rhythm of N, V, N, V, N, V.
Trigeminy
A dominant rhythm of N, N, V,N, N, V, N, N, V.
Tachy
The average heart rate is equal to or greater than the tachycardia
limit.
Brady
The average heart rate is equal to or less than the bradycardia limit.
Pacer Not Pacing
No pace pulse detected for 1.75 x average R-to-R intervals following
a QRS complex (for paced patients only).
Pacer Not Capture
No QRS complex detected for 300 milliseconds following a pace
pulse (for paced patients only).
Missed Beats
No beat detected for 1.75 x average R-R interval for HR <120, or
No beat for 1 second with HR > 120 (for non-paced patients only), or
No beat detected for more than the set pause threshold.
Multiform PVC
Multiform PVCs detected in Multif PVCs Window (which is adjust-
able).
Nonsus V-Tach
The consecutive PVCs < the V-Tach PVCs limit but > 2, and HR ≥ the
V-Tach Rate limit.
Pause
No QRS detected within the set time threshold of pause.
A-Fib
Continuously detect that the RR intervals of normal sinus beats
irregular and no P wave of normal sinus beats.
Irr Rhythm
Consistently irregular rhythm.
Pauses/min
The number of pauses detected per minute.
Arrhythmia message
Description
Category
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...