Measurement Specifications
Product Specifications
A - 14
TMS60 Operator’s Manual
A.7.3
SpO
2
A.7.3.1
Masimo SpO
2
Module
NOTE
•
A functional tester or SpO
2
simulator can be used to determine the pulse
rate accuracy.
SpO
2
Measurement range
1% to 100%
Resolution
1%
Accuracy*
70-100%: ±2.0% (measured without motion)
70-100%: ±3.0% (measured with motion)
1-69%: Not specified
Low Perfusion Conditions
Pulse amplitude: > 0.02%
Light penetration: > 5%
Accuracy: ±2%
Data update rate
≤2s
*SpO
2
accuracy was determined by testing on healthy adult volunteers in the range of 60% to 100% SpO
2
against a laboratory CO-Oximeter. SpO
2
accuracy was determined on 16 neonatal NICU patients ranging in age
from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over
range of 70-100% SaO
2
with a resultant accuracy of 2.9% SpO
2
.
The following table shows the accuracy (A
rms
) for the SpO
2
sensors in four kinds of SaO
2
ranges. For the Fitting
Curve of the SpO
2
sensors, refer to the
"Fitting Curve for Masimo SpO2 Sensors" on page A - 15
.
SaO
2
range
Measured A
rms
Values
LNCS Adtx, LNCS Pdtx
LNCS DCI, LNCS DCIP
70% to 100%
± 2%
2%
70% to 80%
1.55%
0.60%
80% to 90%
1.07%
0.54%
90% to 100%
1.64%
0.60%
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...