Maintenance
Verifying the ECG at the TD60
TMS60 Operator’s Manual
16 - 5
16.6
Verifying the ECG at the TD60
The ECG signal may be inaccurate due to hardware or software problems. As a result, the
ECG wave amplitude becomes greater or smaller. To verify the ECG waveform amplitude:
1.
In the main menu, tap
Parameter Setup
.
2.
Tap
ECG
and set
Filter
to
Monitor
.
3.
In the main menu, tap
Maintenance
.
4.
Input the maintenance passcode and tap
Accept
.
5.
In the
Maintenance
menu, tap
General
.
6.
Enable
Calibrate ECG
.
A square wave appears on the screen and the message
ECG Calibrating
is dis-
played.
7.
Compare the amplitude of the square wave with the wave scale. The difference
should be within 5%. If the difference exceeds 5%, contact your service personnel.
8.
After completing the verification, disable
Calibrate ECG
.
16.7
NIBP Accuracy Test
The NIBP accuracy test is required at least once a year or when you doubt the measured
NIBP result. Contact your service personnel to perform NIBP accuracy test.
16.8
NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at
least once a year or when you doubt the measured NIBP result. Contact your service per-
sonnel to perform NIBP leakage test.
16.9
Nurse Call Test
The nurse call test checks whether this functionality can work properly. Contact your ser-
vice personnel to perform the nurse call test.
16.10 Electrical Safety Tests
The electrical safety tests aim to detect abnormalities that, if undetected, could prove
dangerous to either the patient or the operator. Contact your service personnel to per-
form the electrical safety tests.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...