Product Specifications
Measurement Specifications
TMS60 Operator’s Manual
A - 11
Defibrillation energy absorption
≤ 10% (100 Δ load)
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a
PACE marker:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2 ms
Rise time: 10 μs to 100 μs
Amplitude: ≥ 0.2mV RTI
Pace pulse rejection
When tested in accordance with the IEC60601-2-27: 201.12.1.101.13,
the heart rate meter rejects all pulses meeting the following condi-
tions.
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2 ms
Rise time: 10 μs to 100 μs
No overshoot
About 50% pulse of ANSI/AAMI EC13 (5d) can trigger the pulse
detector, the least changing rate is 20 V/s RTI.
HR
Measurement range
Adult: 15 bpm to 300 bpm
Pediatric: 15 bpm to 350 bpm
Resolution
1 bpm
Accuracy
± 1 bpm or ± 1%, whichever is greater.
Sensitivity
160μV (accuracy: 0.1mV or 25%, whichever is greater; and the abso-
lute amplitude of signal is greater than 0.15mV)
HR averaging method
In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3)
of IEC60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4
most recent RR intervals are averaged to compute the HR. Other-
wise, heart rate is
computed by subtracting the maximum and minimum ones from
the most recent 12 RR intervals and then averaging them.
The HR value displayed on the monitor screen is updated every sec-
ond.
Tall T-wave rejection capability
When the test is performed based on Clause 201.7.9.2.9.101 b) 2)of
IEC60601-2-27, the heart rate meter will reject all 100 ms QRS com-
plexes with less than 1.2 mV of amplitude, and T waves with T-wave
interval of 180 ms and those with Q-T interval of 350 ms.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...