Technical Alarm Messages at the TD60
Troubleshooting
14 - 6
TMS60 Operator’s Manual
14.3
Technical Alarm Messages at the TD60
The following table lists the major technical alarm messages displayed in the message
area of the TD60.
The Alarm Indication column in the table below is capable of three different indication
types: A, B, C.
■
A: The alarm sound will be silenced, the alarm light will continue to indicate the
alarm, a
√
will appear before the alarm message. The
symbol appears on the
top of the main screen.
■
B: The technical alarm will be changed to the prompt message, it will not longer
make sound or be indicated by the alarm light.
■
C: The technical alarms are cleared, there will be no alarm indications.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause
Solution
ECG
ECG Lead
Off
Low*
B
The electrode has
become detached from
the patient or the lead
wire has become discon-
nected from the adapter
cable.
Check the connec-
tions of the elec-
trodes and
leadwires.
ECG XX**
Lead Off
Low*
B
ECG
Module
Error
High
A
An error occurred to the
ECG module.
There is a problem with
the communications
between the module and
the TD60.
Restart the TD60. If
the problem persists,
contact your service
personnel.
ECG
Noise
Low
A
The ECG signal is noisy.
Check for any possi-
ble sources of signal
noise around the
cable and electrode,
and check the
patient for great
motion.
ECG
Cable
Type
Error
Low
B
Connect wrong ECG lead-
wire.
Reconnect the 3 or 5
lead ECG leadwire.
HR
Over-
range
Low
A
HR exceeds the measure-
ment limit.
Contact Mindray or
your service person-
nel.
*
The asterisk (*) means the alarm level is configurable.
**
XX is the ECG lead name.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...