12.3.1.3
Arrhythmia Threshold Settings
When an arrhythmia violates its threshold, an alarm is triggered.When HR is less than 30
bpm, it is recommended to set asystole delay to 10 s.
Arrhythmia event
Range or Option
Default
Step
Unit of
measure
PVCs /min
1 to 100
10
1
minute
Pauses/min
1 to 15
8
1
None
Asystole Delay
3 to 10
4
1
second
Tachy (HR High)
60 to 300
Adult: 120
Pediatric: 160
5
bpm
Brady (HR Low)
Adult: 15 to 115
Pediatric: 15 to 120
Adult: 50
Pediatric: 75
5
bpm
Extreme Tachy
60 to 300
Adult: 160
Pediatric: 180
5
bpm
Extreme Brady
Adult: 15 to 115
Pediatric: 15 to 120
Adult: 35
Pediatric: 50
5
bpm
Multif PVCs Window
3 to 31
15
1
beat
V-Tach Rate
100 to 200
130
5
bpm
V-Tach PVCs
3 to 99
6
1
beat
Pause Threshold
1.5, 2, and 2.5
2
0.5
second
V-Brady Rate
15 to 60
40
5
bpm
V-Brady PVCs
3 to 99
5
1
beat
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...