Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 28
TMS60 Operator’s Manual
12.2.8
SpO
2
Settings
The following table lists the SpO
2
settings in the
SpO
2
tab.
Options
Description
Settings*
Activation State
Configures whether or not to enable the
SpO
2
alarm.
On
, Off
Alarm Priority
Configures the SpO
2
alarm levels.
High,
Med
Record on Alarm
Selects whether or not to activate the
recorder when an SpO
2
alarm is triggered.
On,
Off
Sweep Speed
Configures the SpO
2
waveform sweep
speed.
6.25 mm/s, 12.5 mm/s,
25.0 mm/
s
, 50.0 mm/s
SpO
2
Desat Limit Activa-
tion State
Configures whether or not to enable the
SpO
2
Desat alarm.
On
, Off
Sensitivity
(Masimo only)
The option is not configurable. The option setting is synchronous with the set-
ting at the TD60. See
Configuring the SpO2 Setup" on page 9 - 5.
SpO
2
High Limit (%)
Configures the SpO
2
high alarm limit.
(Low limit + 1) to 100
The default is
100
.
SpO
2
Low Limit (%)
Configures the SpO
2
low alarm limit.
0 to (high limit - 1)
The default is
90
.
PR High Limit (bpm)
Configures the PR high alarm limit.
(Low limit + 2) to 300
The default for adult is
120
.
The default for pediatric is
160
.
PR Low Limit (bpm)
Configures the PR low alarm limit.
18 to (high limit - 2)
The default for adult is
50
.
The default for pediatric is
75
.
SpO
2
Desat Limit (%)
Configures the SpO
2
Desat low limit.
0 to 100
The default is
80
.
Averaging
(Masimo only)
The option is not configurable. The option setting is synchronous with the set-
ting at the TD60. See
Configuring the SpO2 Setup" on page 9 - 5 for details.
*
The factory default settings are in bold.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...