Troubleshooting
Technical Alarm Messages at the TD60
TMS60 Operator’s Manual
14 - 7
Resp
Elec-
trode
Poor
Contact
Prompt
/
The electrode has been
used for a long time or
the electrode contact is
poor.
Check the electrode
application. Reposi-
tion or replace the
electrodes if neces-
sary.
SpO
2
SpO
2
Sensor
Off
Low*
B
The SpO
2
sensor has
become detached from
the patient or the mod-
ule.
There is a fault with the
SpO
2
sensor.
An unspecified SpO
2
sen-
sor has been used.
Check the sensor
application site and
the sensor type, and
make sure the sen-
sor is not damaged.
Reconnect the sen-
sor or use a new sen-
sor.
SpO
2
Sensor
Fault
Low
C
SpO
2
No
Sensor
Low
B
SpO
2
Module
Error
High
A
SpO
2
Too
Much
Light
Low
A
There is too much light
on the SpO
2
sensor.
Move the sensor to a
place with lower
level of ambient
light or cover the
sensor to minimize
the ambient light.
SpO
2
No
Pulse
Low
A
SpO
2
sensor failed to
obtain pulse signal.
Move the sensor to a
site with better per-
fusion.
SpO
2
Unplugg
ed
Low
C
SpO
2
module connector
is disconnected from the
TD60.
Reconnect the SpO
2
module to the TD60.
PR Over-
range
Low
A
The measured PR value
exceeds the measure-
ment range.
Contact Mindray or
your service person-
nel.
NIBP
MPAN
Discon-
nected
Med
C
The MAPN is discon-
nected.
Enable the MPAN
switch.
NIBP
Clock
Needs To
Be Set
Med*
C
The button cell does not
have sufficient charge.
Reset the system
time for the BP10.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause
Solution
*
The asterisk (*) means the alarm level is configurable.
**
XX is the ECG lead name.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...