Introduction
Configuring the TD60
11 - 2
TMS60 Operator’s Manual
11.1
Introduction
The
Maintenance
menu at the TD60 provides access to the system settings such as loca-
tion, device name, alarm settings, quick keys, screen lock, and password updates. Enter-
ing this menu
requires
a passcode.
11.2
Entering the Maintenance menu
1.
In the main menu, tap
Maintenance
.
2.
Input the maintenance passcode.
3.
Tap
Accept
to enter the
Maintenance
menu.
11.3
Configuring the General Menu
Select
General
to configure the display auto off, language, location, notch filter, ECG
lead labeling, SpO
2
module, SpO
2
tone, and enable or disable ECG calibration.
1.
In the
Maintenance
menu, tap
General
.
The current setting displays to the right of the option.
2.
Select the options described in the following table.
Options
Description
Settings*
Display Auto Off
Configures the time for display auto off.
1 min
, 2 min, 5 min, 15 min 30
min, Off
Language
Configures the system language.
ENGLISH
, FRENCH
Location
Changes the hospital and department
names
N/A
Notch Filter
Configures the ECG Notch filter. This
option is used to filter out AC line noise
from the ECG waveform.
50 Hz,
60 Hz
, Off
The default is
50 Hz
when the
monitor is not configured for the
US.
ECG Lead Labeling
Changes the ECG lead labeling.
AHA
, IEC
Calibrate ECG
Enables or disables the ECG verification.
On,
Off
Refer to
for details.
SpO
2
Module
Changes the SpO
2
module.
Masimo
, Nellcor
*
The factory default settings are in bold.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...