Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 12
TMS60 Operator’s Manual
3. Select
Exit
to save the settings and close the
Analysis Lead Setup
menu.
Other Settings
In the
HR
section of the
ECG
tab, select
Other Settings
. The
Other Settings
menu dis-
plays.
The following table lists all settings in the
Other Settings
menu.
12.2.2.3
ECG Display
The ECG digital area displays:
1.
Parameter label
2.
HR value
3.
High HR alarm limit
4.
Low HR alarm limit
5.
Pauses threshold
6.
Pauses per minute value
7.
Pauses per minute label
8.
PVCs threshold
9.
PVCs per minute value
10.
PVCs per minute label
Options
Description
Settings*
Paced
The option is unavailable at the CMS. The paced status can be set at the TD60 only.
Refer to
"Setting the Paced Status" on page 7 - 10
for details.
Pacer Reject
Configures whether or not to reject the pace
pulses.
■
On
: the pace pulses are not counted as
extra QRS complexes.
■
Off
: the pace pulses are not rejected.
On,
Off
This
Pacer Reject
option is only available when
Paced
is set to
Yes
at the TD60.
Filter
Configures the ECG filter in all operating
modes.
■
Monitor
: use under normal
measurement conditions.
■
ST
: use when the ST monitoring is
applied.
Monitor
, ST
*
The factory default settings are in bold.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...