Monitoring at the CMS (Version Earlier than 03.00)
Monitoring with the TMS60 at the CMS
12 - 26
TMS60 Operator’s Manual
■
A clinician has observed clinically questionable arrhythmia calls
■
“Learning” occurred during a Leads Off condition
12.2.5.5
Configuring the QRS Threshold
The minimum detection threshold is approximately 0.16 mV. In case the P waves are very
tall, one might consider moving the minimum QRS detection threshold up to be above
the level of the P waves, so events like ventricular standstill are not missed.
Two horizontal lines (one below and one above the baseline) appear on the screen.
These represent the current minimum detection threshold on the positive and negative
sides of the baseline so manual adjustment for both positive and negative going QRS's
can be made. Using the commands provided, move the minimum detection threshold
up or down to ensure it is above the level of the P waves but below the peak of the R-
wave. Since the P-wave height could vary a little from beat to beat, do not set the hori-
zontal line representing the minimum detection threshold at or barely above the level of
the peak of the P wave. Ensure that it is at least one or two millimeters above the peak of
the P wave but below the peak of the R-wave.
1.
In the
Arrhythmia Analysis
section of the
ECG
tab, select
Minimum QRS Thresh-
old
.
The
Minimum QRS Threshold
menu displays, as shown in the following figure.
Minimum QRS Threshold menu
CAUTION
•
Initiate ECG relearning only during periods of normal rhythm and when
the ECG signal is relatively noise-free. If ECG learning takes place during
ventricular rhythm, the ectopics may be incorrectly learned as the nor-
mal QRS complex. This may result in missed detection of subsequent
events of V-Tach and V-Fib.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...