Direct Control Procedures
User’s Guide
349
Figure 213:Removing the Manifold Plug
6. Click
OK
from the
Remove Manifold Plug
Figure 214:Remove Manifold Plug Dialog
7. To install the capillary array, see
“Removing and Replacing the Capillary Array”
901290L.AI
Guide Pins
Eject Lever
Array Fitting
gel strand
Optics Base Plate
Array Fitting
(bottom)