Direct Control and Replenishment
348
CEQ™ 8000 Genetic Analysis System
Figure 212:Manifold Access Cover
6. Lift the Eject Lever to release the Manifold Plug.
7. Grasp the Manifold Plug tab (Figure 213) and then:
a. Pull the plug approximately one inch out of the manifold.
b. Touch the tip of the plug to the bottom of the Optics Base Plate.
c. Hold and wait five seconds for the gel strand to dry.
d. Pull the plug out and set it aside for future use.
e. Wipe gel strands off of the instrument using a damp tissue.
Manifold
Access
Cover
(removed)
900485e.AI
900487e.AI
900488e.AI
HV
LASER
PWR
900485e .AI
900487e .AI
HV
Capillary Temperature
Control Cover
(open)
Manifold
Access
Cover
captive screw
900488e .AI
Plenum
Assembly
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...