Direct Control Procedures
User’s Guide
337
Figure 201:Manifold Access Cover
6. Loosen the two Plenum Assembly captive screws (Figure 202), pull the Plenum
Assembly straight back and away from the instrument and set it aside.
Slowly remove the Plenum Assembly as the electrode block
may disengage from its mounting posts and become damaged.
Figure 202:Plenum Assembly Removal
7. Lift the Eject Lever (Figure 203) to release the Array Fitting.
Manifold
Access
Cover
(removed)
900485e.AI
900487e.AI
900488e.AI
HV
LASER
PWR
900485e .AI
900487e .AI
HV
Capillary Temperature
Control Cover
(open)
Manifold
Access
Cover
captive screw
900488e .AI
Short Plenum
Assembly
CAUTION
Plenum Assembly
captive screw
captive screw
rubber latch
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...