Fragment Analysis Procedures
228
CEQ™ 8000 Genetic Analysis System
The CEQ System Software is equipped with its own set of default analysis parameters.
These are used to estimate unknown fragment lengths from a calibration curve based
on the migration times of the size standard fragments. The default set of analysis
parameters cannot be edited.
Selecting a Saved Analysis Parameter Set
1. In the
Select Analysis Parameter Set
area of the window, select the
parameter set from the choices available in the drop-down list box.
2. The
Parameter Set Summary
area of the window will display summary
information about the selected analysis parameter set.
These include:
•
The
date and time
the analysis parameter set was created or last modified.
•
The selected
size calibration standard
,
calibration curve
and optional
dye mobility correction
. These parameters will be discussed in the
following sections.
•
The currently assigned
locus tags
that have been selected for this Study.
3. When you have made your selection, click
Next
to proceed to the
Analyze Data
window.
If the available parameter sets are not applicable to your Study, you
will need to create a new set of analysis parameters or edit an existing
set to meet the requirements of your Study. Refer to the “Defining the
Fragment Analysis Parameters” section of this manual or the Online
Help for more information.
Analyzing the Data
With the new data selected and the method of analysis defined, the third step in
creation of the Study is to analyze the data.
The
Analyze Data
window will display all of the samples that have been selected for
analysis, the date each was collected, the status (unanalyzed) of each sample, and the
parameter set being applied to the Study (see the following figure).
1. To begin the analysis, click the
Analyze
button.
2. As the system proceeds with the analysis, the progress bar will be active and the
lower dialog box will record and display the analysis log for each sample as it is
analyzed.
3. To stop the analysis while it is in progress, click the
Stop
button. The system will
stop the analysis when it has finished with the sample in progress.
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...