Viewing the Last Analysis Performed
User’s Guide
147
b. Select the
Save Collected Data
option (if desired).
c. Select the
Perform Shutdown Method
(to purge the manifold and
capillaries and set capillary chamber temperature to 35°C), then select
OK
.
Viewing the Last Analysis Performed
To view the last analysis performed on the CEQ System, perform the following steps.
1. With a sample plate running (and with
Automatic Analysis
selected), select
Tools | View Last Analysis
from the menu.
2. In the
Sample Plate
dialog box, select a sample or multiple samples and then
click
Open
. If the last analysis was a sequence analysis, the Sequence Analysis
module will be launched and the sequence analyzed data displayed.
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...