Direct Control and Replenishment
330
CEQ™ 8000 Genetic Analysis System
Loading the Sample Plate or Buffer Plate
1. Select
Direct Control | Access Plates
from the menu.
Figure 193:Access Plates Dialog
2. Click
Start
. The Capillaries Exposed dialog (Figure 194) will appear.
Figure 194:Capillaries Exposed Dialog
3. Wait until the dialog warning turns green.
4. Open the Sample Access Cover (Figure 190) and lift to the vertical locking
position.
•
If
Sample Plate
•
If
Buffer Plate
Loading the Sample Plate
1. Make sure the Wetting Station is installed or perform the procedure “
” on page 328, and then return here.
2. Align the Sample Plate Guide Pin with the notched corner of the sample plate and
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...