Routine Maintenance
360
CEQ™ 8000 Genetic Analysis System
Bulk Disposal
When performing this procedure, use an exhaust ventilation unit
that meets TLV requirements.
1. Using a 1L side arm flask as a trap, connect the trap to a vacuum.
2. Attached a pipette to the trap using chemical resistant tubing.
3. Aspirate the buffer/gel mixture from each well and dispose of it in a
hazardous liquid organic waste container.
4. After complete removal of all buffer/gel mixture from the plates, allow
the empty plate to sit in the ventilation unit for 24 hours.
5. Dispose of the plate in a solid waste container.
Dispose of buffer/gel mixture in accordance with all applicable
federal, state and local environmental regulations concerning
hazardous liquid waste.
Disposal of the Capillary Array
After removing the expended capillary array from the instrument,
dispose of it in a solid hazardous waste container.
Disposal of the Gel Cartridge
After removing the expended gel cartridge from the instrument,
use a lab spatula or wooden dowel to push any remaining gel into a
liquid hazardous waste container. Dispose of the empty cartridge
in a solid hazardous waste container.
Disposal of D.I. Water/Gel Mixture from the Wetting Tray
When performing this procedure, use an exhaust ventilation unit
that meets TLV requirements.
1. Pour the D.I. water/gel mixture into a hazardous liquid organic waste
container.
2. Rinse the Wetting Tray with water and dispose of the rinse in the liquid
waste container.
WARNING
WARNING
WARNING
WARNING
WARNING
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...