Routine Maintenance
352
CEQ™ 8000 Genetic Analysis System
Cleaning the Capillary Array
The capillary array windows must kept free of any contaminants.
Otherwise, high backgrounds and/or drifting baselines will occur.
All background levels should be below 6000 counts.
Water used during this procedure must be fresh, distilled, deionized
water (18 Mohm/cm water).
1. Select
Replenish | Release Capillary Array
from the Run Module Main
window. Wait for the Remove Capillary Array dialog to appear.
2. Open the Sample Access Cover (Figure 215) and lift to the vertical locking
position.
3. Open the Capillary Access Cover and lift to the vertical locking position.
4. Unlatch the two rubber latches holding the Capillary Temperature Control Cover
and lift to the vertical locking position.
5. Loosen the Manifold Access Cover captive screw (Figure 216), remove the cover
and set it aside.
Figure 216:Manifold Access Cover
6. Lift the Eject Lever (Figure 217) to release the Array Fitting.
7. Grasp the Array Fitting tab (Figure 217) and then:
a. Pull the fitting approximately one inch out of the manifold.
b. Touch the tip of the fitting to the bottom of the Optics Base Plate.
c. Hold and wait five seconds for the gel strand to dry.
CAUTION
Manifold
Access
Cover
(removed)
900485e.AI
900487e.AI
900488e.AI
HV
LASER
PWR
900485e .AI
900487e .AI
HV
Capillary Temperature
Control Cover
(open)
Manifold
Access
Cover
captive screw
900488e .AI
Plenum
Assembly
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...