Analyzing Sequence Data
User’s Guide
157
Heterozygote Detection Tab
Select the
Heterozygote Detection
tab as shown in Figure 95 to define the
parameters the system will use to detect heterozygotes. These parameters will only be
used if the
Detect Heterozygote
option is selected in the
General
tab.
Figure 95: Heterozygote Detection Tab
Detect Heterozygotes
When this item is checked, the analysis will include heterozygote detection based on
the remaining parameters specified in the
Heterozygote Detection
tab.
When the
Detect Heterozygotes
check box is checked, the
Call Threshold
is
disabled (grayed) but the current Call Threshold value remains in the text box. The
system will use a value of zero for the Call Threshold and will not call “N’s” when
attempting to detect heterozygotes.
% of Average Peak Spacing
Use this parameter to determine the range around each peak that the system will search
for a possible heterozygous peak. The system will determine the local peak to peak
spacing using an average of the spacing for a 10 base range around the base of interest.
The system will then find the specified percent (e.g., 70%) of the local peak to peak
spacing and look in that range from the center of each peak for a possible
heterozygous peak. For instance, if a value of 70% is entered, the system will look
35% of the average peak spacing to the right and 35% to the left for additional peaks.
The range is 0 to 100% and the default is 70%.
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...