Run Module
User’s Guide
39
When stopping a direct control process:
Stop Options -
Specifies what process to stop:
Stop All
cancels all the currently running
processes.
Stop High Voltage Only
stops the high
voltage but continues to inject the sample.
Monitor
Baseline
Baseline monitoring displays the baseline data trace. To
ensure that the optics are working correctly and that the
capillaries are clean, view the baseline when a new
capillary array is installed, after the capillaries have been
filled with gel for the first time and an optical alignment
has been performed. The baseline trace should be low and
relatively flat (less than 6000 counts).
If
Autosave
is selected from the dialog box, baseline
data will be stored under the Sample Data folder (for the
selected project) in the Data Manager. If Monitor Baseline
is not ended manually, baseline data will be collected until
a sample plate is run or until six hours have elapsed from
the start of the baseline monitoring process.
Diagnostics
Used to view the instrument version, laser power, PC
Settings, monitor status, and to home the plates and/or the
gel pump.
Reset
Used to re-synchronize system hardware with the
firmware and to reset the system log.
Item
Description
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...