User’s Guide
281
Starting a New AFLP Analysis
When you have finished setting the analysis parameters, click
OK
to generate the
New AFLP Analysis
.
Figure 155:New AFLP Analysis window
Viewing the Cluster Analysis
When the clustering operation is completed, a table will be displayed showing the
results of the AFLP analysis. The format of this table can be displayed as either rows
or columns of data for ease of export to a third party software platform for further
analysis (see sample axis below).
•
The
Bin Column / Row
displays the bin number. Bins with A and B suffixes
will be assigned if two adjacent bins fall in the same integer after rounding of the
Bin X Mean.
•
The
Dye Column / Row
displays the dye included.
•
The
Samples Column / Row
displays the sample count within the bin.
•
The
Fragments Column / Row
displays the fragment count within the bin.
•
The
X-Min Column / Row
displays the minimum fragment size within the bin.
•
The
X-Max Column / Row
displays the maximum fragment size within the bin.
•
The
X-Mean Column / Row
displays the mean fragment size within the bin.
•
The
X-Var Column / Row
displays the fragment size variation within the bin.
•
The
Y-Mean Column / Row
displays the mean peak height within the bin.
•
The
numbered columns / rows
to the right / bottom of the table show the
number of sample components (peaks/fragments) within each bin for each sample.
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...