Routine Maintenance
User’s Guide
357
No more than one 96-well plate should be processed, for each
wetting tray, without replenishing the Wetting Tray.
Periodically check the liquid level in the wetting tray. Liquid level
should NEVER be allowed to rise into the eight recesses of the
wetting tray lid, nor drop below the fill level indicator line (9 mL
minimum). The top surface of the wetting tray lid must remain clean
and dry under any and all circumstances.
Installing the Wetting Tray
1. Rotate the wetting tray retainers outward.
2. Insert the Wetting Tray into the receptacle between the Sample and Buffer plates
(Figure 221) and then gently press it down into the well.
3. Rotate the Wetting Tray Retainers inwards to lock the Wetting Tray in place.
Figure 221:Replacing the Wetting Tray
4. Close the Sample Access Cover and then click the
Done
button of the
Replace
Wetting Tray
If a sample plate is installed in the system, it must be removed
before installing the wetting tray to allow movement of the wetting
tray retainers.
CAUTION
901002L.A
Wetting Tray Well
(front)
Sample Plate
Holder
Wetting Tray
Wetting Tray Lid
Wetting Tray Retainers
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...