Direct Control and Replenishment
328
CEQ™ 8000 Genetic Analysis System
Accessing the Direct Control Window
From the CEQ Toolbar, click on the
RUN
icon. Verify that the Run window is
displayed and then click on the
Direct Control
tab.
Direct Control Procedures
Replacing the Wetting Tray
Removing the Wetting Tray
1. Select
Replenish | Replace Wetting Tray
from the Run Module main menu.
2. Press
OK
when prompted to verify this action.
3. Open the Sample Access Cover (Figure 190) and lift to the vertical locking
position.
4. Remove the sample plate and set aside.
5. Rotate the Wetting Tray Retainers outwards to release the Wetting Tray.
6. Lift the Wetting Tray vertically.
Installing the Wetting Tray
1. Select
Replenish | Replace Wetting Tray
from the Run Module main menu.
2. Press
OK
when prompted to verify this action.
3. Open the Sample Access Cover (Figure 190) and lift to the vertical locking
position.
4. Rotate the Wetting Tray Retainers outward. Insert the Wetting Tray into the
receptacle between the Sample and Buffer plates (Figure 191) and then gently
press it down into the well.
5. Rotate the Wetting Tray Retainers inwards to lock the Wetting Tray in place.
If a sample plate is installed in the system, it must be removed before
installing the wetting tray to allow movement of the wetting tray
retainers.
Содержание CEQ 8000
Страница 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Страница 98: ...84 CEQ 8000 Genetic Analysis System...
Страница 110: ...96 CEQ 8000 Genetic Analysis System...
Страница 120: ...106 CEQ 8000 Genetic Analysis System...
Страница 128: ...114 CEQ 8000 Genetic Analysis System...
Страница 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Страница 154: ...140 CEQ 8000 Genetic Analysis System...
Страница 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Страница 208: ...Sequence Analysis Procedures 194 CEQ 8000 Genetic Analysis System Figure 119 Sequence Results Report...
Страница 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Страница 303: ...User s Guide 289 More information on these and other display modifications can be found in the Online Help...
Страница 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Страница 329: ...Exporting Database Items User s Guide 315 Figure 180...
Страница 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Страница 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...