User Manual
California Instruments
INITIAL SETUP
The initial setup menu can be used to determine the AC source
settings at power up. CAUTION: The initial setup can be used to
power up the AC source with the output on and a high voltage
present at the output. For normal situations, this is not
recommended due to the potential danger to operators. It is
recommended that the initial voltage be set low and/or the
output relay be programmed to OFF for most situations.
LIMIT SETUP
The Limit menu shows the frequency, voltage and current limit
capabilities of the AC source. Any attempt to program the output
beyond these limits will result in a “-222 Data Out of Range
error”. Note that these limits are hardware determined and
cannot be changed by the user. They are shown for reference
only.
UTILITY 2
CONFIGURATION
The Configuration menu shows the installed options. This
screen is for reference only and no fields can be changed by the
user.
ELAPSED TIME
The elapsed time screen, when selected from the UTILITY
menu, will appear for about 3 seconds. The elapsed time shown
is the cumulative amount of time the power source has been on
from its initial build. This value is read only and cannot be
changed by the user.
The same screen also displays the internal AC source ambient
temperature in degrees C.
VIEWING ANGLE
The viewing angle can be used to change the contrast ratio of
the LCD display. The range of the viewing angle parameter is
from -10 to +10. Setting the right viewing angle is matter of
personal taste. Set this parameter to a value that is most
comfortable for the user.
NSG 1007 Series II
93
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...