User Manual
California Instruments
9.4.5.4Output Wiring
The EOS-1 or EOS-3 is wired in series with the output of specified Teseq power systems. The
EOS terminal block marked SOURCE should be wired to the output of the power source/system.
The terminal block marked LOAD is wired to the user's load. Refer to
Figure 9-112 and Figure 9-113 in this section for examples of proper connections. Refer to
Section 3.5 to determine the appropriate wire gauge needed for the output wiring.
Note in Figure 9-112 the sequence of instruments. This is the recommended sequence between
the iX source and the equipment under test. For systems that do not include all the instruments
shown, maintain the sequence of the remaining instruments.
Note also that for all applications the remote sense wiring must be connected
before
the EOS
unit.
Note: The output voltage of the iX/i-series power sources into the EOS box may
be at hazardous potentials as high as 300 volts line to neutral (600 volts line to line).
Wiring used between the power sources and the EOS and between the EOS and the
user’s load must be insulated to withstand this potential.
Note: The SOURCE and LOAD terminals of the EOS will be at the hazardous live
potentials of the iX/i-series power sources driving it even if EOS box has not been
switched on.
NSG 1007 Series II
201
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...