User Manual
California Instruments
3.6.6I/O Option – J58
This connector is reserved for control of the EOS option. Do not connect anything else to this
connector.
3.6.7Function Strobe / Trigger Out – J22-31 / J22-14
A function strobe output is available on the System Interface connector. This open collector output
may be used to trigger external equipment when voltage or frequency change occurs on the AC
source.
This output generates a low-going pulse, > 400
µ
s in duration, that indicates voltage or frequency
change. Since this is an isolated output, an external DC supply and pull-up resistor, 22K
Ω
, 1/8 W is
required. To create a TTL level output, a +5V or +3.3V DC supply is required. To create a signal for
viewing on a scope, a higher DC voltage such as a 9V battery may be used.
Connect the DC supply and pull-up resistor as shown.
+ 5 Vdc
J22-31
J22-14
Function Strobe
High
Function Strobe
Common
Centronics
Connector
System
Interface
+5
0
> 400 uS
Figure 3-13: Function Strobe Connection.
When running list transients on the AC source, the LIST:TTLTrigger SCPI command may be used
reassign the operation of the Function Strobe output as a trigger output. An output pulse is
generated for each logic “1” in the TTLTrigger list. See the NSG 1007 Programming Manual P/N
7000-982 for details on the transient list system.
Trigger BNC Breakout Box
A convenient trigger BNC breakout box is available from Teseq which facilitates connection to the
Function Strobe / Trigger Output signal on the system interface connector. A 9V DC battery is used
to provide the required DC supply. This small box can be ordered through customer service or sales
under P/N 7000-481-1 (for NSG 1007-3 and NSG 1007-5) or P/N 7000-481-2 (for all multi-chassis
i/iX systems).
NSG 1007 Series II
43
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...