4.2.6TRANSIENTS Menu
Figure 4-39: TRANSIENTS menu
The transient menu provides access to the transient list data. The iM Series does not support
transient programming. Both the i and the NSG 1007 II have a transient list of up to 100 data
points. This is represented by 100 transient step numbers from 0 through 99. From the Transient
menu, the desired transient step type can be selected. Based on the user’s choice, the relevant
transient type sub menu will be shown. The START/EDIT SEQUENCE sub menu allows the user
to review and change any transient step or execute the transient list. When executing a transient
list, transient steps are executed in a ascending numerical order. Steps that are not defined are
skipped.
The following entries can be found in the TRANSIENTS menu:
Entry
Description
VOLT SURGE/SAG
Voltage surges and sags are temporary changes in amplitude.
The output voltage will change from its present value to a user
specified value for a specified duration. (Sag if the value is
lower, surge if the value is higher.) After this period has expired,
the output voltage returns to a user specified end value. This
value may or may not be the same as the value present prior to
the start of the sag or surge.
VOLT SWEEP/STEP
Voltage sweeps cause the output voltage to change from the
present value to a user specified end value at a specified rate of
change. A voltage step on the other hand is an instantaneous
change in output voltage. The new value will be held for the
duration period specified by the user. The final output voltage
value of a sweep and a step transient step should be different
than the value at the start of the transient step or no change in
output value will occur.
FREQ SWEEP/STEP
This transient type is similar to a voltage sweep/step except it
affects the frequency. Refer to the previous paragraph.
VOLT/FREQ SWEEP/STEP
This transient type combines the previous two types into a single
step. The effect is that of changing the output voltage and
frequency simultaneously.
Note: While this transient is programmed as a single transient
step, two list entries are required to store this information. As
such, every VOLT/FREQ SWEEP/STEP used will consume two
list entries at a time.
START/VIEW SEQUENCE
This entry allows the user to switch to the transient execution
menu. This menu provides a list of all available transient list
steps and their sequence numbers. From this menu, transient
list execution can be started.
The same menu can be used to view or edit any available
transient list step or erase a step using the backspace key.
NSG 1007 Series II
79
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...