California Instruments
POWER INTERRUPT
This test requires a numeric entry value equal to the test number. The tests are grouped as follows:
•
Test numbers 1 through 15 are for all Standard and Groups. See Figure 9-86 for details of the tests.
•
Test numbers 16 and 17 for all equipment that does not incorporate digital circuit. Test number 16 will
drop the output to zero voltage for 50 ms. Test number 17 will drop the output to zero voltage for 200
ms. Test number 18 is used for DC equipment and will drop the output for 1 sec.
•
Test numbers 21 through 26 are applicable for Groups 2 and 3 only. Output frequency will be set to the
F1 value for 1 second prior to the test. The output frequency will remain set to the F2 value when the
test is completed. This will allow the user to apply sequence of power interrupts. See Figure 9-87 for
detail of the tests.
V (NOM)
% of V NOMINAL
(V MIN)
0 VOLTS
T1
T2
T3
DO160 Table 16-1: Test conditions for equipment with digital circuits.
NOTES
1:
Definitions:
T1
Power interrupt time
T2
Time it would take for the applied voltage to decay from V (nom) to zero volts.
T3
Time it would take for the applied voltage to rise from zero to V (nom) volts.
V MIN
The minimum level (expressed as a percentage of V NOMINAL) to which the applied
voltage is permitted to decay.
2:
Tolerance to T1, T2, T3 = ± 10%
3:
Test condition numbers 8 and 15 are for category Z, dc powered equipment only.
Applicable
Category:
A
A, Z
Z
A, B, Z
A, Z
Z
Test
Condition
No.
1**
2
3
4
5
6
7
8
9
10
11
12
13
14
15
T1 (ms)
2**
10
25
50
75
100
200
1000
10
25
50
75
100
200
1000
T2 (ms)
<1
20*
20
20
20
20
20
20
50*
50*
50
50
50
50
50
T3 (ms)
<1
5
5
5
5
5
5
5
20
20
20
20
20
20
20
%V Nom.
(V min)
0
50
15
10
5
0
0
0
80
50
0
15
5
0
0
*
Voltage will not reach zero in this test condition.
**
Equipment performance standards may require to repeat test n°1 with T1 varying from 5 to 200 ms by step
defined in the test equipment performance standards (step typically comprised between 5 ms and 20 ms depending
on equipment design.
Figure 9-86: Power Interrupt
166
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...