User Manual
California Instruments
RUN SINGLE
RUN SINGLE command will run the test once. The Variation test is defined by the REDUCE TO,
FALL TIME, HOLD TIME and RISE TIME parameters. These parameters must be set before
starting the test. The following is a description of these parameters.
REDUCE
TO:
The lowest voltage level as a percentage of the nominal voltage. Thus, 0% is 0
Volts. 100% is full nominal voltage.
FALL TIME:
The time in seconds it will take the output to reach the REDUCE TO voltage.
Values must be entered in seconds. A 0.000 sec time may be used for abrupt
voltage drops instead of voltage sweeps to support Edition 2.0 of the test
standard. This requires firmware 2.38 or higher however.
HOLD TIME:
The time in seconds the output will hold at the REDUCE TO voltage. Values
must be entered in seconds.
RISE TIME:
The time in seconds the output will reach the NOMINAL voltage from the
REDUCE TO voltage. Values must be entered in seconds.
The timing of voltage variations is different between editions 1.0 and 2.0 of the IEC 61000-4-11
test standard. While the original standard used voltage ramps on both sides of the variation test
level, the Edition 2.0 standard calls out an abrupt drop to the test level, followed by a voltage
ramp back to the nominal voltage. Thus, for Edition 2.0, the fall time must be set to 0.000
seconds.
Furthermore, where under Edition 1.0 rise and fall times were specified in seconds, they are
specified in number of cycles in Edition 2.0. This is an effort to normalize both 50 Hz and 60 Hz
voltage variations.
Refer to Figure 9-97 and Figure 9-98 for an illustration of how these parameters affect the V
RMS output under the different standard revisions.
Figure 9-97: EN 61000-4-11 Voltage Variation specification- Edition 1.0
Figure 9-98: EN 61000-4-11 Voltage Variation specification- Edition 2.0
182
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...