6.Calibration
The Routine Calibration should be performed every 12 months. Non-routine Calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
All standard models and configurations of the iX /i / iM Series may be calibrated using a PC
running Windows 2000/XP, and the latest version of the CIGuiSII AC source control software.
Refer to the CIGuiSII online help file of the CIGUISII software for additional procedures and
guidance. To download the latest version of the CIGUISII, visit
For iM Series II model calibration, refer to the I Series II model information in this chapter.
6.1Calibration Equipment
Digital Phase Meter:
Krohn-Hite, Model 6620 or equivalent. (0.01° resolution, 0.02°
accuracy.)
Digital Multimeter:
Fluke 8506A or equivalent.
HP 34401A
10 milliohm Current Shunt:
Isotek Model RUG-Z-R010-0.1 or equivalent
1 milliohm Current Shunt:
Isotek Model RUG-Z-R001-0.1
(10001i/iX, 15001i/ix and
NSG 1007-15-208-MODE):
Load Bank:
Various high power load resistors will be needed.
Computer (Optional)
PC to operate Teseq GUI.
6.2Calibration Screen Access
Output
To access the OUTPUT CALIBRATION screen for the first time press the MENU key twice to
display the MENU 3 screen. Press the
↑
or
↓
key several times to highlight OUTPUT CAL.
Press the ENTER key.
Type 5000 and press the ENTER key to show the OUTPUT CALIBRATION screen.
Press the PHASE key to select the phase to be calibrated for a 9003i/iX ,10002i/iX or 15003i/iX
power system.
Measurements
To show the MEASUREMENT CALIBRATION screen follow the steps in paragraph 6.2.
Select the MEASUREMENT CAL function instead of OUTPUT CAL from the MENU 2 screen.
If another CALIBRATION screen has been accessed since power-up no password is needed. If
prompted for a password, use the value 5000.
NSG 1007 Series II
139
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...