User Manual
California Instruments
4.6.3.3Trigger delay
The trigger delay field allows the user the set the amount of pre- or post-trigger data that should
be used when positioning the data acquisition window with respect to the trigger moment.
POST-TRIGGER DELAY
A positive trigger delay value means the acquisition window is delayed by the amount of time
specified. In this case, the actual trigger moment itself is no longer present in the acquisition
buffer. This situation is shown in Figure 4-66 where a 20 ms trigger delay is used after triggering
on phase A = 180°. The fundamental frequency of the output is 50 Hz. The trigger point is
indicated by the dashed line. It occurs on the first 180 degree point that occurs after the user
presses the ENTER key while on the START field. Once the trigger occurs, the acquisition holds
off the specified 20 ms at which point the data requested is captured. Using a positive trigger
delay value always yields post trigger data.
Figure 4-66: Positive trigger delay (Post trigger data)
Positive trigger delay values may be set from 0.0 ms to 1000.0 ms (1 second) in 0.1 ms
increments. The value may be entered directly from the keyboard or using the knob.
120
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...