Front Panel Entry
While it is possible to perform IEC 61000-4-13 testing from the front panel of the NSG 1007 AC
power source, it is recommended to use the provided GUI Windows program for report
generation. This also provides a more convenient way to perform Class 1 and User class tests
as test levels can be saved to disk.
To perform a test from the keyboard, select the APPLICATIONS screen from the MENU 2
screen. The APPLICATIONS screen will appear as shown in Figure 9-100.
Figure 9-100: Application menu
Scroll to the IEC 61000-4-13 entry using the up and down cursor keys. Press the ENTER key to
select the IEC 61000-4-13 main menu. The screen will appear as shown in Figure 9-101.
Figure 9-101: IEC 61000-4-13 Menu
9.3.3.1IEC 61000-4-13 Screen Parameters
The IEC 61000-4-13 screen has the following common fields for all IEC 61000-4-13 test groups.
STATE
The STATE field must be set to ON before the test can run. Initial setup conditions must be met
to run the test.
RUN
The RUN field has the following selections:
1. ALL
Run all IEC 61000-4-13.
2. GROUP
Run the selected group.
3. RANGE
Run the selected range for the selected group.
4. POINT
Run the selected frequency for the selected range and selected group.
Note: The RANGE and POINT do not apply to all groups. See individual groups for details.
When the test is running, this field will have the following selection:
1. ABORT
The test will stop running and all parameters will return to the state prior to the
start of the test.
2. PAUSE
The test will pause at the time the ENTER key is pressed while the PAUSE field
is selected. The field will change to RESUME. When the ENTER key is
pressed again, the test will resume from the point it stopped.
NSG 1007 Series II
185
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...