User Manual
California Instruments
6.3Routine Measurement Calibration
Note: The Fluke 8506A Digital Multimeter must be used for the following calibration. The
8506A must be set to the AC HI ACCUR mode for all AC measurements.
Connect the test equipment to the power source as shown in Figure 6-2. If the power system is
a multi-phase system with one controller, the DVM for calibrating the measurement voltage
should always be connected to the Remote Sense connector (TB3) on the Phase A power
source.
The shunt must be connected to the power source as shown in Figure 6-78. If the Current
measurement can’t be successfully performed, adjust the Current Measurement Pot. This
adjustment is described in the Non-routine Calibration section of this manual. If the DC current
measurement displays more than 70 counts on the display, perform the non-routine current
monitor adjustment.
Connect the load to the output. Recommend current shunt values are 0.010 ohm for
3001/5001/9003/15003 and 0.001 ohm for 10001, 15001.
To calibrate all measurement functions, the desired value for the measurement value of current
or voltage must be entered for the corresponding calibration value. Make the indicated
adjustments by typing in the desired display value. This should be the value indicated by the
external DVM. If a 10 milliohm current shunt is used for current, 300 mV represents 30 amps.
The Measurement Calibration Table is a summary of the measurement calibration procedure.
The following text is a detailed explanation of the procedure.
PARAMETER
POWER SYSTEM
3001 / 9003
5001 / 15003
10001 / 30003
15001
AC Current Full-Scale
6.0
Ω
, 3 KW
3.6
Ω
, 5 KW
1.8
Ω
, 10 KW
1.3
Ω
, 15 KW
DC Current Full-Scale
13.5
Ω
, 1.5 KW
4.8
Ω
, 3.5 KW
2.4
Ω
, 7 KW
1.6
Ω
, 10.5 KW
Table 6-9: Calibration Load For Each Phase
140
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...