User Manual
California Instruments
SWEEP GROUP
If the SWEEP group is selected, the screen will appear as shown in Figure 9-104. The screen
has the following parameters that are unique to the group:
Figure 9-104: IEC 61000-4-13 Sweep
1. LEVEL
Sets the percentage level of the inter harmonics relative to the fundamental.
The level is fixed for the entire frequency range, which is defined by
FRANGE. To change level, the test state must be off and user class must be
selected.
2. FRANGE
The range is selected by rotating the shuttle. The range selection is required
to change the level for the selected range. Also, an individual range sweep
is possible by selecting the desired range and selecting RUN RANGE from
the RUN field.
3. STEP
The step size defines the inter harmonics sweep points. Step size is fixed for
the entire sweep range. The STEP size can be changed when the test state
is off.
4. IHFREQ
This field will show the current inter harmonic frequency when the test is
running. An inter harmonic frequency can be selected using the shuttle. To
run a frequency without sweeping through the entire range, a RUN POINT
can be selected.
5. RESONANT This field is used to report the resonant points, if any, after running the sweep
test. No editing is allowed in this field. When the test is completed, a display
of harmonics current versus frequency plot is available. To view the graph,
select the field and press the ENTER key.
188
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...