User Manual
California Instruments
4.7Transient Programming
4.7.1Introduction
Transient programming provides a precise timing control over output voltage and frequency
changes. This mode of operation can be used to test a product for susceptibility to common AC
line conditions such as surges, sags, brownouts and spikes. By combining transient
programming with custom waveforms [], virtually any AC condition can be simulated on the
output of the AC source.
The default voltage mode is FIXED which means the output voltage is constant and remains at
the level set by the user. Changes made to the output voltage made from the PROGRAM 1
menu take effect immediately. In front panel operation mode, the voltage and frequency slew
rates (rate of change) are always at their maximum of 1E9 V/s and 1E9 Hz/s. Slew rate
programming is only possible over the IEEE-488 or RS232C bus. On power up, the AC source
always reverts to the maximum slew rate for both voltage and frequency.
4.7.2Using Transient Modes
The voltage can be programmed in the following transient operating modes:
STEP
causes the output to permanently change to its triggered value.
PULSE
causes the output to change to its triggered value for a specific time, as
determined by the Pulse menu parameters.
LIST
causes the output to sequence through a number of values, as determined by
points entered in the List menu.
FIXED
disables transient operation for the selected function.
4.7.3Step Transients
Step transients let you specify an alternate or triggered voltage level that the AC source will apply
to the output when it receives a trigger. Because the default transient voltage level is zero volts,
you must first enter a triggered voltage before you can trigger the AC source to change the
output amplitude. Step transients can only be programmed through the bus, not the front panel.
Refer to the SCPI Programming Manual for more information about programming Step transients
and triggers.
122
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...