User Manual
California Instruments
7.3.5Unit Shuts Down after 1-2 Seconds
CAUSE
SOLUTION
Output shorted
Remove output short
Output grossly overloaded.
Remove overload.
Operating load with too high inrush or start
up currents.
Consult factory for application advice.
7.3.6No Output and No Lights on Front Panel
CAUSE
SOLUTION
Input circuit breaker switched off.
Switch the breaker on.
No input power to TB3.
Ensure 3 phase power is getting to TB3.
Unit tripped on overvoltage or overcurrent.
Turn circuit breaker off - wait five seconds
- turn circuit breaker back on.
7.3.7No Output But “Power On” LED on Front Panel is Lit
CAUSE
SOLUTION
“OUTPUT ON” switch is turned off.
Turn OUTPUT ON switch to “ON”.
REMOTE SHUTDOWN logic line at J22
pin 36 is shorted to D COM or A COM.
Remove connection from J22 pin 3.
Current limit programmed down or to zero.
Program current limit higher.
Voltage programmed down or to zero.
Turn amplitude control up.
Unit tripped on overvoltage or overcurrent.
Turn circuit breaker off - wait five seconds
- turn circuit breaker back on.
152
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...