User Manual
California Instruments
VOLT SURGE
This test will produce voltage surge defined by Table 9-33. This test will be repeated three times
with ten seconds intervals. The voltage values are halved for 14.0V category B equipment.
Category
Surge 1
Surge 2
Volt
Dwell(msec)
Volt
Dwell(msec)
A
46.3
100
37.8
1000
B
60
100
40
1000
Z
80
100
48
1000
Table 9-33: Abnormal Voltage Surge
9.1.1.6.3Emergency Test
The Emergency test is selected from the DO-160 DC Main Menu. This test will set the output
voltage to 18V for 28V equipment and to 9V for 14V equipment. The test will last for 30 minutes.
The
←
key (backspace) will terminate the test at any time.
NSG 1007 Series II
175
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...