User Manual
California Instruments
Figure 9-141: WH-Meter Screen With Function Active
Note
: Changing from stop to start will stop the measurement and will maintain the last data
record for the watt-hour meter. To restart the measurements, the field is toggled to the stop
position from the start position and the previous data will be reset to zeros.
WATT HR, POWER and PK CURR will display the data for the phase selected with the phase
key. If the phase is selected for all phases in a two or three phase system, the WATT HR and
POWER fields will show the sum of all phases. PK CURR will remain at the last selected phase.
Note
: Changing Mode of operation from single phase to three phases while the Watt-hour meter
is in the active state will generate an error: “Input buffer full”. This error will halt the
measurements. To clear the error, the Watt-hour meter must be switched to the stopped state.
It is advised that the Watt-hour meter should be set to the stop state before changing the mode
of operation.
250
i Series II / NSG 1007 II / iM Series II
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...