3.6.8Remote Inhibit – J22-36 / J22-27
The Remote Inhibit input J22 pin 36 can be used to open and close the output relay of the AC
power source. This input overrides the state of the output relay programmed from the front panel
or the bus. It may be used for safety interlock purposes.
The default level for remote inhibit is a logic low or contact closure between pin J22-36 and pin J22-
27 (D COM). This will cause the output voltage to be programmed to 0.0 volts and the output relays
to open. Alternative, the level can be reversed using the “output:ri:level high” command over the
bus.
The mode of operation of the remote inhibit can be changed using the Output mode bus command.
This command selects the mode of operation of the Remote Inhibit protection. The following
modes can be selected:
LATChing
A TTL low at the RI input latches the output in the protection shutdown state,
which can only be cleared by OUTPut:PROTection:CLEar.
LIVE
The output state follows the state of the RI input. A TTL low at the RI input turns
the output off; a TTL high turns the output on.
OFF
The instrument ignores the RI input.
The RI output state is saved at power down. The factory default state is LIVE.
For details on programming the remote inhibit function, refer to the iX programming manual (P/N
7000-982).
NSG 1007 Series II
45
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...