User Manual
California Instruments
9.3.3.4INTERHARMONICS
A single inter harmonic frequency may be generated using the INTERHARMONICS screen. This
screen allows insertion of any inter harmonic from 1Hz to 2400Hz in 1Hz steps. The amplitude
level of the harmonics range is from 0 to 20% of the programmed voltage. To select the inter
harmonics screen, press the menu screen until the MENU 2 screen appears as shown in Figure
9-110. Select INTERHARMONICS using the up or down key.
Figure 9-110: MENU 2 SCREEN
INTERHARMONICS SCREEN
The inter harmonics screen will appear as shown in Figure 9-111. This screen has the following
parameters:
Figure 9-111: INTERHARMONICS SCREEN
1. REFERENCE
This field will enable the Inter harmonics generator if ON is selected and
will disable the inter harmonics generation if OFF is selected.
2. REF. COUPL
If turned on, the REF. VOLT will follow the programmed voltage value.
1. REF. VOLT
This field must be programmed to a value if REF. COUPL is off. This
value has a range from 0 to 230 volts. This value is set to a value
equal to the phase A voltage if the REF. COUPL is turned on.
2. VOLTAGE
This field programs the Inter harmonic level in percentage of the
REF.VOLT. This value has a range from 0 to 20%
NSG 1007 Series II
195
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...