User Manual
California Instruments
9.2.6.2VOLTAGE VARIATION TESTS
From the IEC1000-4-11 Main Menu screen shown in Figure 9-94, scroll to the VOLTAGE
VARIATIONS entry using the up and down cursor keys. Press the ENTER key to select the
VOLTAGE VARIATIONS menu. The screen will appear as shown in Figure 9-96.
Figure 9-96: Voltage Variation screen
RUN ALL
The RUN ALL selection will cause the following automated test sequence suggested by the
standard to be run:
Step
Type
Start V in % of
U
T
Dwell time
End V in % of
U
T
Delay between
steps (s)
1
VSweep
100
2
40
0
2
Hold
40
1
40
0
3
VSweep
40
2
100
10
4
VSweep
100
2
40
0
5
Hold
40
1
40
0
6
VSweep
40
2
100
10
7
VSweep
100
2
40
0
8
Hold
40
1
40
0
9
VSweep
40
2
100
10
10
VSweep
100
2
0
0
11
Hold
0
1
0
0
12
VSweep
0
2
100
10
13
VSweep
100
2
0
0
14
Hold
0
1
0
0
15
VSweep
0
2
100
10
16
VSweep
100
2
0
0
17
Hold
0
1
0
0
18
VSweep
0
2
100
10
Table 9-37: Voltage Variations Test Performed During RUN ALL
The user can change the NOMINAL Ut voltage for this test. The RUN ALL Command line will
change to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate
the test at any time and the output voltage will return to the nominal value.
NSG 1007 Series II
181
Содержание 10002iX
Страница 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Страница 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Страница 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Страница 59: ...User Manual California Instruments Figure 3 26 Connection With MODE Option NSG 1007 Series II 59 ...
Страница 131: ...User Manual California Instruments Figure 5 73 Power Source Module Block Diagram NSG 1007 Series II 131 ...
Страница 134: ...User Manual California Instruments Figure 5 74 5001i Internal Layout 134 i Series II NSG 1007 II iM Series II ...
Страница 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Страница 164: ...User Manual California Instruments Figure 9 84 Voltage Modulation 164 i Series II NSG 1007 II iM Series II ...
Страница 204: ...User Manual California Instruments Figure 9 114 NSG 1007 15 208 CTS EOS3 LR3 204 i Series II NSG 1007 II iM Series II ...
Страница 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...
Страница 223: ...User Manual California Instruments Figure 9 118 Example Connections With INA 2151 NSG 1007 Series II 223 ...