CPS100 SERIES OPERATION MANUAL
SECTION 6:
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual
Page 89 of 231
Buzzer Setting:
The buzzer can be turned on or off by
pressing the “Buzzer” soft key and toggling
the setting between ON and OFF.
When ON, a sound will sound each time the
user presses a key on the front panel or
presses the shuttle.
If an audible indication is not desired, turn
the buzzer sound OFF.
Power On State Setting:
The Power On state (P/O State) soft key allows the user to select the programmed state of
the power source after turning on the input power. Available selections are:
•
OFF
The power source will NOT store any setting when powered off and initialize
with factory default at power on.
•
LAST The power source will store the last setting in effect when powered off and
recall this same setting at power on. That means the unit will power up in
the same conditions as when it was turned off.
•
USER In this mode, a user-defined setup can be assigned for recall when the
power source is turned on. This means the power source will always turn on
with this specific setup.
Example:
To set a power on setting of AC+DC coupled output, Vac = 100V, F = 50Hz and
Vdc = 20V, use the following steps:
1.
Press the P/O State soft key and scroll to the USER entry, then press [Enter] to
confirm. Available setting parameters are now visible in the lower portion of the
LCD.
2.
Move the cursor to the “Couple = “, press [Enter] to confirm.
3.
Rotate the shuttle to highlight the “AC+DC” option, press [Enter] to confirm.
4.
Use the [0] ~ [9] keys and [Enter] to set the Vac, F and Vdc fields to 100, 50 and 20
respectively.